1DOC272.53.80033User instructionsGel filled ORP/Redox electrode: Model MTC10101, MTC10103, MTC10105, MTC10110, MTC10115 or MTC10130Safety InformationP
10 DOC272.53.80033Advanced operationParameter-specific settings can be changed through the Full Access Options menu. Details about menu navigation, av
DOC272.53.80033 11.6. If prompted, enter a name for the new method settings. Additional changes made to the settings of an existing method are automat
12 DOC272.53.800336. If prompted, enter a name for the new method settings. Additional changes made to the settings of an existing method are automati
DOC272.53.80033 13TroubleshootingMessage or symptomPossible cause ActionProbe not supportedSoftware not updatedUpdate the HQd software to the newest r
14 DOC272.53.80033Drifting/Inaccurate readingsContaminated platinum disc Clean the electrode.Clogged referenceRinse reference junction holes with deio
DOC272.53.80033 15PartsDescription Quantity Catalog NumberIntelliCAL ORP/Redox gel-filled electrode, standard, with 1 m cable1 MTC10101IntelliCAL ORP/
HACH Company World HeadquartersP.O. Box 389Loveland, Colorado 80539-0389 U.S.A. International customers:Tel (800) -227-4224 (U.S.A. only) Tel +001 (97
2 DOC272.53.80033Preparation for use1. Remove the locking ring and shroud (rugged model).2. Before first use, remove the tape from the reference junct
DOC272.53.80033 3CalibrationCalibration notes• Probes are initially calibrated at the factory. However, regular calibration by the user is recommended
4 DOC272.53.80033Note: Procedures also apply for rugged model probes.1. Connect the electrode to the meter.2. Add fresh ORP/Redox standard to beaker o
DOC272.53.80033 5Run check standardThe run check standard feature validates instrument performance between sample measurements. Use the run check stan
6 DOC272.53.80033Note: Procedures also apply for rugged model probes.1. Prepare the electrode for use (Preparation for use on page 2).2. Connect the e
DOC272.53.80033 7Measurement - conversion to SHE referenceFor some applications it is customary to report redox potential readings relative to the sta
8 DOC272.53.800331. Fill a 10 ml burette with a standard titrant solution with a normality that is 5-10 times that of the sample. 2. Connect the elect
DOC272.53.80033 9MaintenanceClean the probe if the response time becomes slow (Clean the probe on page 9).StorageNote: For optimal performance, the re