DOC272.53.80021
User instructions
Luminescent Dissolved Oxygen Probe: Model LDO10101,
LDO10103, LDO10105, LDO10110, LDO10115 or LDO10130
Safety Information
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed.
Overview
Figure 1 on page 2 shows the LDO101 series probe, a luminescent dissolved oxygen (LDO
®
)
probe. The LDO10105, LDO10110, LDO10115 or LDO10130 rugged probe is available with a
5, 10, 15, or 30 meter cable. The LDO10101 or LDO10103 standard probe is available with a
1 or 3 meter cable. The probe provides measurement of the dissolved oxygen concentration in
wastewater, drinking water and general applications. For BOD applications, use the
LBOD10101 probe equipped with LDO technology and an incorporated stirrer.
Specifications
Specifications are subject to change without notice.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC),
European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no
charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return
end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
Specification Details
Dissolved oxygen range 0.1 to 20.0 mg/L (ppm) 1 to 200% saturation
Dissolved oxygen accuracy
± 0.1 mg/L for 0 to 8 mg/L
± 0.2 mg/L for greater than 8 mg/L
% saturation resolution 0.1%
Temperature range 0 to 50 °C (122 °F)
Storage temperature 0 to 40 °C (104 °F)
Response time T90% at 10 sec (when stirred)
Temperature resolution 0.1 °C (32.18 °F)
Temperature accuracy ± 0.3 °C (32.54 °F)
Pressure resolution 1 hPa
Pressure accuracy ± 0.8%
Minimum sample depth 25 mm (0.984 in.)
Dimensions
Standard diameter: 15 mm (0.59 in.), Length: 103 mm (4.1 in.), Total length (standard):
200 mm (7.9 in.), Total length (rugged): 220 mm (8.7 in.), Cable length: 1 or 3 meter (3.3
or 9.8 ft.). Rugged diameter: 15 mm (0.6 in.), Length: 16 mm (0.63 in.), Cable length: 5,
10 15 or 30 meter (16.4, 32.8, 49.2, or 98.4 ft.)
Cable connection Digital output and connector compatible with HQd meters
Warranty Probe is covered by a 3 year warranty. Sensor cap is covered by a 1 year warranty.
Vue de la page 0
1 2 3 4 5 6 ... 13 14

Résumé du contenu

Page 1 - User instructions

1DOC272.53.80021User instructionsLuminescent Dissolved Oxygen Probe: Model LDO10101, LDO10103, LDO10105, LDO10110, LDO10115 or LDO10130Safety Informat

Page 2 - Preparation for use

10 DOC272.53.800217. If prompted, enter a name for the new method settings. Additional changes made to the settings of an existing method are automati

Page 3 - Calibration

DOC272.53.80021 116. If prompted, enter a name for the new method settings. Additional changes made to the settings of an existing method are automati

Page 4 - METHODS key, and

12 DOC272.53.80021Slope out of range Probe not prepared for sampleAllow probe to reach equilibrium in a water-saturated air environment and repeat cal

Page 5 - Measurement

DOC272.53.80021 13Replacement partsDescriptionQuantity Item NumberIntelliCAL LDO probe, standard, with 1 m cable 1 LDO10101IntelliCAL LDO probe, stand

Page 6 - Maintenance

HACH Company World HeadquartersP.O. Box 389Loveland, Colorado 80539-0389 U.S.A. International customers:Tel (800) -227-4224 (U.S.A. only) Tel +001 (97

Page 7 - Figure 3 Sensor exploded view

2 DOC272.53.80021Preparation for use1. Verify that the LDO sensor cap and iButton are installed correctly. Avoid touching the sensor cap with a hand,

Page 8 - Advanced operation

DOC272.53.80021 3CalibrationWater-saturated air (100%) calibrationCalibration notes• Probes are initially calibrated at the factory. However, regular

Page 9

4 DOC272.53.80021Note: Procedures also apply for rugged model probes.1. Connect the probe to the meter. The LDO101 probe contains a factory calibratio

Page 10 - Change calibration options

DOC272.53.80021 5MeasurementNote: Procedures also apply for rugged model probes.Auto salinity correction (optional for HQ40d meter)Dissolved substance

Page 11 - Troubleshooting

6 DOC272.53.800212. Push METER OPTIONS and select the CDC401 Settings.3. Select Current Method then Hach Salinity. Select OK.Note: To change additiona

Page 12 - Troubleshooting (continued)

DOC272.53.80021 7Figure 3 Sensor exploded view1 Shroud 7 Cap seal2 Locking ridges (8x) 8 Temperature sensor3 Locking ring 9 Locking groove4 Sensor cap

Page 13 - Accessories

8 DOC272.53.80021To remove the shroud1. Unscrew the locking ring. 2. Slide the shroud and locking ring off the probe.To install the shroud1. Place the

Page 14 - P.O. Box 389

DOC272.53.80021 9Change measurement optionsMethods are groups of default or user defined settings relevant to specific applications. If the meter is s

Commentaires sur ces manuels

Pas de commentaire