DOC024.52.93023ORBISPHERE Model K-M1100 Sensor and Model410 AnalyzerUser Manual12/2013, Edition 9
8General Information1.4 Product recycling informationENGLISHElectrical equipment marked with this symbol may not be disposed of in European public dis
98Maintenance and Troubleshooting14.2.3 Sensor spot replacement3. When unscrewed completely simply lift out the old sensor spot. Pull off the maintena
99Maintenance and Troubleshooting14.3 Storage, handling and transportationProtect the instrument against the elements: rain, splashing, direct sunligh
100Maintenance and TroubleshootingNote: Use the Diagnostic view for troubleshooting (see Sensor diagnostics on page 95).Note: If the RTC battery is ru
101Maintenance and Troubleshooting14.5 List of events and alarmsTable 4 List of EventsEvent type Name DescriptionBit mask value (32 bits long) Infor
102Maintenance and Troubleshooting
103Section 15 Accessories and Spare Parts15.1 Instrument15.2 SensorImportant Note: The shelf-life of the K1100/M1100 LDO spot is 2 years prior to firs
104Accessories and Spare Parts15.3 Insertion and sampling devices15.4 K1100 kits to order (KTO’s)Part No. Description29006.0 EPDM O-ring set for flow
105Accessories and Spare Parts15.5 M1100 kits to order (KTO’s)K1100-KTO-W Kit contains K1100-S00 sensor, 410/K/W1C00000 instrument, 3m cableK1100-KTO-
106Accessories and Spare PartsDGKM1100-W1125 Kit contains M1100-S00 sensor, 410/M/W1C10000 instrument, 5m cableDGKM1100-W2011 Kit contains M1100-S10 s
107Accessories and Spare Parts15.6 Instrument optionsA number of differently configured 410 instruments are available for use with the either the K110
9General InformationSVENSKAElektronikutrustning som är märkt med denna symbol kanske inte kan lämnas in på europeiska offentliga sopstationer efter 20
108Accessories and Spare Parts
109Section 16 Glossary16.1 Gas units16.2 Generic terms and definitionsUnit Meaning % airpercentage, by weight. A concentration of 100% air corresponds
110GlossaryParallel communicationParallel communication represents a connection in a computer system in which the bits of a byte are transmitted over
HACH COMPANY World HeadquartersP.O. Box 389, Loveland, CO 80539-0389 U.S.A.Tel. (970) 669-3050(800) 227-4224 (U.S.A. only)Fax (970) 669-2932orders@hac
10General Information1.5 Product disposalNote: The following only applies to European customers.Hach Lange is committed to ensuring that the risk of a
11General Information1.6 Restriction of hazardous substances (RoHS)The European Union RoHS Directive and subsequent regulations introduced in member s
12General Information
13Section 2 Specifications2.1 Sensor descriptionsORBISPHERE K1100 sensorThe high accuracy K1100 sensor is a luminescent sensor measuring dissolved oxy
14SpecificationsSensor Sample temperatureMeasurement from -5 to 50°C (23 to 122°F)Sensor resistant to temperature from -5 to 100°C (23 to 212°F)Sample
15Specifications2.3 General principle of operationOptical sensing of oxygen originates from the work of Kautsky in 1939 where he demonstrated that oxy
16SpecificationsThe oxygen partial pressure (pO2) is then linked to the corresponding phase-shift measurement (Φ) to build the sensor calibration curv
17Specifications2.5 Security level tableA cross means that the user who has this user security level can access this function or setting (see User man
18Specifications2.6 Default parametersThe table below indicates the factory default configurations. The instrument has these settings when started for
19Specifications2.7 M1100 EHEDG Certification
20Specifications
21Section 3 InstallationThis section provides necessary information to install and connect the analyzer. The installation of the analyzer should be pe
22Installation3.2 Installation check listFor a complete installation, proceed to the following actions following the instructions in this manual very
23Installation3.3 Wall mount and pipe mount instruments3.3.1 Instrument dimensionsFigure 4 Wall/Pipe mount instrument dimensions (in millimeters)
24Installation3.3.2 Wall mounting3.3.3 Pipe mountingAttach the U bracket (provided) to the wall with two screws (not provided).Tilt the instrument sli
25Installation3.3.4 Connections (bottom of instrument)A square key is provided to open the instrument front panel locks. The two locks are located on
26Installation3.4 Panel mount instrument3.4.1 Instrument dimensionsFigure 8 Panel mount instrument dimensions (in millimeters)
27Installation3.4.2 MountingFigure 9 Panel mount bracket frame1. Cut an opening in the panel to accommodate the bracket frame provided (this is the s
1Table of ContentsSection 1 General Information...
28Installation3.4.3 Connections (bottom of instrument)Note: The type of power cable connection will vary depending on the instrument specification.Alt
29Installation3.5 Connectors assembly instructions3.5.1 Cable gland wiring instructionsA waterproof cable gland is provided each time a cable must be
30Installation3.5.2 USB-B client adapter cable3.6 Connection to mains power supply3.6.1 Power supply connection (low voltage instruments)For low volta
31Installation3.6.2 Power supply connection (high voltage instruments)High voltage instruments (100-240 VAC) have a 4-pin male connector pre-wired int
32InstallationWire the female connector as follows:1. Take the narrow end of the connector (4) in one hand and the main body (2) in the other and unsc
33Installation3.7 Connections to electronic boardsNote: Any loose connection wires should be bundled tightly together with the use of nylon cable ties
34Installation3.7.3 Measurement board3.8 Measurement alarm relaysFigure 17 Measurement boardFigure 18 Connector J7Connector J7 (inputs & outputs
35Installation3.9 Sensor installation3.9.1 Sensor positioning information The sensor must be installed in a socket or flow chamber that allows contact
36Installation3.9.4 Sensor removal• If not using the ORBISPHERE 32003 insertion/extraction valve (see details in The 32003 insertion/extraction valve
37Installation3.9.6 The 32003 insertion/extraction valveThe ORBISPHERE 32003 insertion/extraction valve (illustrated below) allows for sensor removal
2Table of Contents3.10.2 ORBISPHERE flow chambers...39Section 4
38Installation3.9.9 Instrument connectionsThe sensor is supplied with a cable having a LEMO connector at both ends. One end is attached to the sensor
39InstallationThe gas bottle is not supplied and must be purchased locally. To ensure the calibration works correctly, the calibration gas bottles mus
40Installation
41Section 4 User Interface4.1 InstrumentThe instrument front panel provides these user interfaces:• Touch screen acting as display, touch pad and keyb
42User Interface4.2.1 Function keys on the header barShortcut to the user login window. Pressing this button for more than 2 seconds calls the ID and
43User Interface4.2.2 Menu navigation4.2.3 Rolling list4.2.4 Virtual keyboardFigure 27 Main menu windowPressing the “menu” button in the header bar c
44User Interface4.2.5 Identification and authorization level Note: To get to level 0, press the unlock button and OK, without entering any ID or passw
45User Interface4.3 Main menu structureThis is the structure of the main menu which is used to control every functionality of the instrument. These su
46User Interface
47Section 5 View MenuFigure 33 View menu
3Table of Contents10.2 Configure security ...
48View Menu5.1 Selection of the view styleNumeric viewThis is the default view. Display shows the numeric measurement value identified for the gas mea
49View Menu5.2 Configuration of the view stylesSkewnessnegativepositiveFigure 34 SkewnessAn asymmetric frequency distribution is skewed to the left i
50View Menu
51Section 6 Measurement Menu6.1 Instrument configurationFigure 36 Measurement menuContinuous mode descriptionContinuous mode is typically used for pr
52Measurement Menu6.2 Measurement configurationMeasurement configuration• Sensor: Sensor model.•Medium: Liquid or gas phase.• Gas unit type: Partial,
53Measurement MenuMeasurement alarms configurationSet the thresholds for the low/high concentration levels, according to the application. Each alarm t
54Measurement Menu6.3 Measured data storageOxygen interference configurationThese options are available to take into account the influence of some com
55Section 7 Calibration MenuCalibrations can only be performed once the instrument has been installed, configured, and the channel has been set up. Yo
56Calibration Menu7.1 Sensor calibrationSensor calibrationThe sensor can be calibrated manually on an ad hoc basis. By default, the mode is set to zer
57Calibration Menu7.2 Calibration configurationCalibration configurationThis option can be invoked directly from the main calibration menu by selectin
4Table of Contents
58Calibration Menu7.3 Zero calibration7.4 100% humid air calibration (high range sensors only)7.5 High level adjustment7.6 Sensor verificationZero cal
59Calibration Menu7.7 Barometric pressure calibration7.8 Calibration reportsBarometric pressure calibrationNote: The barometric sensor has been factor
60Calibration Menu
61Section 8 Inputs/Outputs MenuFigure 39 Inputs/Outputs menu
62Inputs/Outputs Menu8.1 Configure snooze8.2 View inputs/outputs8.3 RelaysConfigure snoozeIn the event of an alarm, the “snooze” button stops the inst
63Inputs/Outputs Menu8.4 Analog outputsFigure 40 Analog outputs menu
64Inputs/Outputs MenuAnalog outputsThere are three analog outputs available. These outputs are configurable in terms of function, content, and behavio
65Inputs/Outputs MenuThe following table lists the default configuration. The first two events on the list are pre-set and only the priority can be ch
66Inputs/Outputs MenuAnalog outputs (continued)Calibration of the analog outputThe calibration of the analog output is aimed at aligning the internall
67Inputs/Outputs Menu8.5 Analog output characteristicsAnalog output characteristics“Linear” analog output The "Linear" output is the default
5Section 1 General Information1.1 DisclaimerThe information in this manual has been carefully checked and is believed to be accurate. However, Hach La
68Inputs/Outputs MenuAnalog output characteristics (continued)“Tri-linear” analog outputThe "Tri-linear" output brings benefits over the “Li
69Inputs/Outputs MenuTri-linear Range Measurement M Resolution R4-20 mA1: AOL > I > 4 M=MLL+(ML-MLL)*(I-4)/(AOL-4) R=(ML-MLL)*20/((AOL-4)*1010)
70Inputs/Outputs Menu
71Section 9 Communication MenuFigure 41 Communication menuThe external RS-485 port of the main board is directly connected to a RS-485 bus (single tw
72Communication Menu9.1 RS-485 simple mode configurationThis protocol allows the instrument to output data to an external device (PLC, SCADA, PC, etc.
73Communication Menu9.1.1 Data availableAll individual data are separated by at least one tabulation character (ASCII code=0x09).For the cyclic measur
74Communication MenuGas sensor calibration report exampleCalibration report nb 1Calibration mode . . . . . . . . . . . .Manual high level adjustmentDa
75Communication MenuMeasurement file example 6 measurements are illustrated below:Nr mm/dd hh:mm:ss Gas Temp Mask Barom Ext.P Ph
76Communication Menu9.1.2 Example of useIn this example we use: • One PC with a RS232 port.• One "RS-485<->RS232 converter"Procedure:1
77Communication Menu9.2 PROFIBUS-DP communication (optional)9.2.1 InstallationOn the ORBISPHERE CD, there is an “Orbi3218.gsd” and an “Orbi3218.bmp” f
6General Information1.3.2 Safety precautions• The power cord plug connection is also used as a main power switch.• The instrument must be connected to
78Communication Menu9.2.2 Input/Output dataThe main board:• Writes the latest measurement data to the Profibus Input Buffer.• Checks if a command writ
79Communication MenuThe gas, temperature and barometric pressure unit values are coded as defined in the following tables:Note: If the instrument stop
80Communication MenuCommandsThe “Command Output Buffer” is formatted as follows:The “Command Input Buffer” is located just after the measurement data
81Communication MenuActivate sensor command - outputActivate sensor command - input9.3 USB-A port (host)This option allows the export or import of dat
82Communication Menu9.4 HTTP/TCP-IP9.4.1 OverviewWhen activated this option downloads data from the instrument directly to a web page that can be acce
83Communication MenuOnce a valid username/password combination has been entered, the initial web page will be displayed giving a list of options:Click
84Communication Menu9.5 Data file transfer through the USB-B port (client)The USB port allows you to copy data files from the instrument to a PC using
85Communication MenuUploaded FilesThere are a number of files that are uploaded to the PC during this process. However, only the text files (with a .t
86Communication Menu9.6 PrinterThis menu provides the facility to print a number of reports directly to a printer. The printer must be connected to th
87Section 10 Security MenuNote: When the instrument is started for the very first time, security is disabled by default. It is highly recommended that
7General Information1.3.4 Precautionary labelsRead all labels and tags attached to the instrument. Personal injury or damage to the instrument could o
88Security Menu10.2 Configure security10.3 User management10.4 User action log fileConfigure securityThis enables defining the users with their access
89Section 11 Products MenuFigure 43 Products menu
90Products MenuProductsThis option allows users to save and/or use previously saved product configurations. A maximum of 100 different product configu
91Section 12 Global Configuration MenuFigure 44 Global configuration menuGlobal configurationThe global configuration option allows users to save, an
92Global Configuration Menu
93Section 13 Services MenuFigure 45 Services menu - part 1
94Services MenuFigure 46 Services menu - part 2
95Services Menu13.1 Sensor diagnostics13.2 Language selection13.3 ClockSensor diagnosticsCalibration timer The instrument can automatically remind the
96Services Menu13.4 Screen13.5 Buzzer13.6 Boards info13.7 Batteries13.8 Software download13.9 End applicationScreenScreen calibration This screen allo
97Section 14 Maintenance and Troubleshooting14.1 Instrument maintenance14.2 Sensor maintenanceThe sensor spot needs to be replaced about once a year,
Commentaires sur ces manuels