DOC023.52.90050TSS Portable handheld measurement instrument for turbidity/solidsUser Manual06/2012, Edition 3A© HACH-LANGE GmbH, 2004–2008, 2012. All

Page 2

WARNINGWARNING10Installation3.2 Powered via rechargeable batteriesThe instrument is operated by six rechargeable NiMH batteries (Figure 2).3.2.1 Inser

Page 3 - Table of Contents

Figure 2 Inserting the rechargeable batteriesWARNING11Installation3.2.2 Install the charger1. Push the slide switch (item 2, Figure 3) on the back of

Page 4

Figure 3 Replacing the adapter12Installation3.2.3 Charging the rechargeable batteriesNote: The instrument must be switched off while charging so that

Page 5 - Section 1 Specifications

Figure 4 Charging the rechargeable batteries13Installation3.3 Connecting the probeRemove the protective cap and connect the probe plug to the instrum

Page 6 - Specifications

Figure 5 Connecting the probe plug to the instrument14Installation3.4 Switching the instrument on and off1. Fully charge the rechargeable batteries (

Page 7 - Section 2 General Information

Figure 6 Description of the keypad15Section 4 Start Up4.1 System start overview1. Switch on the measurement instrument (section 3.4, page 14).2. Conn

Page 8 - 2.5 Probe

Table 1 Navigating in the menuNavigation key DescriptionNavigation key RIGHT/LEFTNavigation key UP/DOWNENTER/ONCLEAR/OFFVertical arrowsCursor> (arr

Page 9 - Section 3 Installation

17Start Up4.2.2.1 Querying internal informationInformation about the probe and instrument can be queried directly from the main menu using the navigat

Page 10 - Installation

18Start Up4.5 Setting the displayIt is possible to custom-define the first two rows of the display. In the standard setting, the first line shows the

Page 11 - 3.2.2 Install the charger

19Start Up4.7 CalibrationFor turbidity measurement, a standard curve C-TU is already stored in the probe. Calibration is not required. However, for so

Page 13 - 3.3 Connecting the probe

20Start UpCalibrate curve C-DS1:7. Select MEMORY and confirm with ENTER/ON.8. Lower the probe into a container that contains a homogeneous sample.9. S

Page 14 - 3.4.1 Register a new probe

21Start Up4.8.2.2 Practical example for turbidity measurement A standard curve C-TU is stored for the turbidity measurement. 1. Connect probe to instr

Page 16 - 4.2.2 Display

Figure 8 Homogeneity in %23Section 5 Operation5.1 CalibrationTurbidity measurements do not need to be calibrated, as a standardized calibration curve

Page 17 - 4.3 Select language

24Operation5.1.2 CalibratingThe instrument can save one turbidity curve and up to four solids calibration curves. This makes measurements in media wit

Page 18 - 4.6 Setting the units

25Operation14. Enter the laboratory values using the navigation keys and confirm with ENTER/ON.• Navigation key LEFT/RIGHT: Jump to the next/previous

Page 19 - 4.7 Calibration

26Operation5.1.3 Manual calibration value correction1. Select MENU, confirm with ENTER/ON. 2. Select CALIBRATE, confirm with ENTER/ON. The instrument

Page 20

27Operation5.2.1 Selecting the calibration curveBefore starting the measurement, the calibration curve that corresponds to the measurement point must

Page 21

28Operation5.5 Delete stored data for all calibration curves1. Select MENU, confirm with ENTER/ON.2. Select SYSTEM, confirm with ENTER/ON. 3. Enter th

Page 22

CAUTION29Section 6 MaintenanceThe cleanliness of the measurement windows in the sensor head is critical to the accuracy of the measurement results!Cle

Page 23 - Section 5 Operation

3Table of ContentsSection 1 Specifications ...

Page 25

Problem/display screen Caused by ActionCommunication with probeEEPROM errorOverflow areaClean the probeOffset errorCalibration errorCalibration insuff

Page 26 - 5.2 Measure

32Troubleshooting

Page 27 - 5.2.4 Stop measurement

33Section 8 Replacement parts and accessories 8.1 Replacement PartsDescription Order-No.TSS Portable handheld instrument(TSS Portable handheld measure

Page 28

34Replacement parts and accessories

Page 29 - Section 6 Maintenance

35Section 9 Contact information HACH CompanyWorld HeadquartersP.O. Box 389Loveland, Colorado80539-0389 U.S.A.Tel (800) 227-HACH(800) -227-4224(U

Page 30 - Maintenance

36Contact informationHACH LANGE D.O.O.Fajfarjeva 15SI-1230 DomžaleTel. +386 (0)59 051 000Fax +386 (0)59 051 010[email protected]www.hach-lange.s

Page 31 - Section 7 Troubleshooting

37Section 10 Warranty and liabilityThe manufacturer warrants that the supplied product is free of material and manufacturing defects and undertakes to

Page 32 - Troubleshooting

38Warranty and liability

Page 33 - 8.1 Replacement Parts

4Section 5 Operation ...235

Page 34

5Section 1 SpecificationsThese are subject to change without notice.Performance specificationsWavelength 860 nm Parameter Turbidity Solids (dry matter

Page 35

6Specifications

Page 36 - Contact information

DANGER7Section 2 General Information2.1 Safety informationPlease read this entire manual carefully before unpacking, setting up, or operating this equ

Page 37

8General Information2.2 Overview of productThe TSS Portable is a handheld measurement instrument for the analytical determination of turbidity and sol

Page 38 - Warranty and liability

9Section 3 Installation WARNING 3.1 Product contentsAfter taking the instrument out of the case, check each part for damage. All the components on th