1DOC272.53.80025User instructionsBOD LDO® Probe: Model LBOD10101Safety InformationRead all labels and tags attached to the instrument. Personal injury
10 DOC272.53.80025TroubleshootingMessage or symptom Possible cause ActionProbe not supportedSoftware not updatedUpdate the HQd software to the newest
DOC272.53.80025 11Calibration failed: outside of acceptance criteria/Temperature out of range/Offset out of limitsWater Saturated air equilibration no
HACH Company World HeadquartersP.O. Box 389Loveland, Colorado 80539-0389 U.S.A. International customers:Tel (800) -227-4224 (U.S.A. only) Tel +001 (97
2 DOC272.53.80025Preparation for use1. Verify that the sensor cap and iButton are installed correctly (Figure 1). Avoid touching the sensor cap with a
DOC272.53.80025 3CalibrationWater-saturated air (100%) calibrationWater-saturated air (100%) calibration notes:• Probes are initially calibrated at th
4 DOC272.53.80025Zero point calibrationZero point calibration notes:• The zero point calibration is only necessary for dissolved oxygen measurements b
DOC272.53.80025 5MeasurementMeasurement notes: • Make sure that the BOD bottle is filled with sample and no air bubbles are present. • The LBOD10101 m
6 DOC272.53.80025MaintenanceGeneral maintenance• Rinse the probe with water to clean. The stirrer does not need to be removed.• Do not scrub the senso
DOC272.53.80025 7Important Note: Avoid handling the black face of the sensor cap. Do not use alcohol or other organic solvents to clean the black face
8 DOC272.53.80025with this name to distinguish them from the default method settings, which cannot be changed. A saved method can be used instead of r
DOC272.53.80025 97. If prompted, enter a name for the new method settings. Additional changes made to the settings of an existing method are automatic