DOC272.53.80025
User instructions
BOD LDO
®
Probe: Model LBOD10101
Safety Information
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed.
Overview
Figure 1 on page 2 shows the LBOD10101 probe, a luminescent dissolved oxygen (LDO
®
)
sensor with an integrated stirring system. The LBOD10101 measures the dissolved oxygen for
BOD (Biochemical Oxygen Demand) determination in BOD bottles. The probe uses the
luminescent dissolved oxygen (LDO) technology (no membranes) and features a replaceable
self-stirring assembly and a 1.8 meter cable.
Specifications
Specifications subject to change without notice.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC),
European electrical equipment users must now return old or end-of life equipment to the Producer for disposal at no
charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return
end-of-life equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
Specification Details
% saturation resolution 0.1%
% saturation accuracy ± 0.59% of reading
Cable connection Digital output and connector compatible with HQd meters
Dimensions
Diameter: 12 mm (0.47 in.), Length: 103 mm (4.06 in.), Total length: 220 mm (8.66 in.),
Cable length: 1.8 meter (5.91 ft)
Dissolved oxygen accuracy
± 0.05 mg/L for 0.0 to 10 mg/L
± 0.1 mg/L for greater than 10 mg/L
Dissolved oxygen range 0.05 to 20.0 mg/L (ppm) 1 to 200% saturation
Minimum sample depth 10 mm
Pressure accuracy ± 3 mbar (0.3 kPa)
Pressure resolution 5 mbar (0.5 kPa)
Response time T90% at 10 s (when stirred)
Temperature accuracy ± 0.3 °C (± 0.54 °F)
Storage temperature 5 to 40 °C (41 to 104 °F)
Temperature resolution 0.1 °C (0.18 °F)
Temperature range 0 to 50 °C (32 to 122 °F)
Warranty
Probe and sensor cap are covered by a 1 year warranty. Use outside of the laboratory
environment may void warranty.
Vue de la page 0
1 2 3 4 5 6 ... 11 12

Résumé du contenu

Page 1 - User instructions

1DOC272.53.80025User instructionsBOD LDO® Probe: Model LBOD10101Safety InformationRead all labels and tags attached to the instrument. Personal injury

Page 2 - Preparation for use

10 DOC272.53.80025TroubleshootingMessage or symptom Possible cause ActionProbe not supportedSoftware not updatedUpdate the HQd software to the newest

Page 3 - Calibration

DOC272.53.80025 11Calibration failed: outside of acceptance criteria/Temperature out of range/Offset out of limitsWater Saturated air equilibration no

Page 4 - Zero point calibration

HACH Company World HeadquartersP.O. Box 389Loveland, Colorado 80539-0389 U.S.A. International customers:Tel (800) -227-4224 (U.S.A. only) Tel +001 (97

Page 5 - Measurement

2 DOC272.53.80025Preparation for use1. Verify that the sensor cap and iButton are installed correctly (Figure 1). Avoid touching the sensor cap with a

Page 6 - Maintenance

DOC272.53.80025 3CalibrationWater-saturated air (100%) calibrationWater-saturated air (100%) calibration notes:• Probes are initially calibrated at th

Page 7 - Advanced operation

4 DOC272.53.80025Zero point calibrationZero point calibration notes:• The zero point calibration is only necessary for dissolved oxygen measurements b

Page 8

DOC272.53.80025 5MeasurementMeasurement notes: • Make sure that the BOD bottle is filled with sample and no air bubbles are present. • The LBOD10101 m

Page 9 - Change calibration options

6 DOC272.53.80025MaintenanceGeneral maintenance• Rinse the probe with water to clean. The stirrer does not need to be removed.• Do not scrub the senso

Page 10 - Troubleshooting

DOC272.53.80025 7Important Note: Avoid handling the black face of the sensor cap. Do not use alcohol or other organic solvents to clean the black face

Page 11 - Troubleshooting (continued)

8 DOC272.53.80025with this name to distinguish them from the default method settings, which cannot be changed. A saved method can be used instead of r

Page 12 - Consumables (continued)

DOC272.53.80025 97. If prompted, enter a name for the new method settings. Additional changes made to the settings of an existing method are automatic

Commentaires sur ces manuels

Pas de commentaire