1DOC272.53.80022User instructionsConductivity probe: Model CDC40101, CDC40103, CDC40105, CDC40110, CDC40115, CDC40130Safety InformationPrecautionary L
10 DOC272.53.80022Replacement partsDescription Quantity Item NumberConductivity probe, standard with 1 m cable 1 CDC40101 Conductivity probe, standard
DOC272.53.80022 11180 µS/cm at 25 °C 100 mL 23075421000 µS/cm at 25 °C 100 mL 144004218.00 mS/cm at 25 °C 100 mL 23074421990 µS/cm at 25 ºC 100 mL 210
HACH Company World HeadquartersP.O. Box 389Loveland, Colorado 80539-0389 U.S.A. International customers:Tel (800) -227-4224 (U.S.A. only) Tel +001 (97
2 DOC272.53.80022CalibrationFigure 1 Conductivity probe overview1 Temperature sensor and 4-pole graphite design conductivity cell3 Rugged probe (5, 10
DOC272.53.80022 3Calibration Notes• Probes are initially calibrated at the factory. However, regular calibration by the user is recommended for the be
4 DOC272.53.80022MeasurementMeasurement notes• Data is stored automatically in the data log when Press to Read or Interval is selected in the Measurem
DOC272.53.80022 5Run check standardThe run check standard feature validates instrument performance between sample measurements. Use the run check stan
6 DOC272.53.80022Clean the probe• For general contaminants, rinse the electrode with deionized water and gently wipe dry with a clean cloth.• For grea
DOC272.53.80022 71. Make sure a probe is connected to the meter.2. Push the METER OPTIONS key and select (CDC401) Settings.3. Select Modify Current Se
8 DOC272.53.800226. If prompted, enter a name for the new method settings. Additional changes made to the settings of an existing method are automatic
DOC272.53.80022 96. If prompted, enter a name for the new method settings. Additional changes made to the settings of an existing method are automatic