Page 1 - NISE sc probe

DOC023.52.90137AN-ISE sc probe AISE sc probe NISE sc probeUSER MANUAL10/2012, Edition 4A© HACH-LANGE GmbH, 2010, 2011, 2012. All rights reserved. Prin

Page 2

10General information2.3.2 AISE sc probeThe AISE sc probe uses the ion-selective electrode technology to measure ammonium ions (NH4+) in a waste water

Page 3 - Table of contents

NOTICE11Section 3 Installation3.1 Unpack the probeRemove the probe from the shipping container and inspect it for damage. Verify that all items listed

Page 4

12InstallationFigure 5 Sensor cartridge packaging 1 Sensor cartridge packaging 6 Socket head screws2 Storage container for sensor cartridge 7 Black

Page 5 - Section 1 Technical data

13Installation3.2.1 Assemble the storage container including sensor cartridgeFigure 6 Storage container for the sensor cartridge 23451Note: Keep item

Page 6 - 1.1 Dimensions

14Installation3.2.2 Remove the cartridge from the storage containerFigure 7 Open the storage container NOTICENote: This black gasket is not required

Page 7 - Section 2 General information

15InstallationFigure 8 Knock the air out of the sensor cartridge NOTICENOTICE3.3 Probe assembly1. Place the black gasket Figure 10, Page 17 in the re

Page 8

16InstallationFigure 9 Sensor cartridge NOTICE3. Align the marker hole on the sensor cartridge with the marker hole on the probe adapter (refer to Fi

Page 9 - 2.3 Functional principle

17InstallationFigure 10 Probe assembly 3.4 Installation of the cleaning unit (optional)Refer to the installation instructions for the cleaning unit f

Page 10 - 2.3.3 NISE sc probe

NOTICE18Installation3.5 Installation of the probe in the sample flowThere are mounts with separate installation instructions available for installatio

Page 11 - Section 3 Installation

19Installation3.5.2 Example of mounting the probeFigure 12 Example of mounting the probe using a rail mount 3.6 Connect the probe to the sc controlle

Page 13

20InstallationFigure 13 Connecting the probe to the sc controller with the screw fitting Number Description Cable color1 +12 VDC Brown 2 Earth Black

Page 14

21Section 4 Operation4.1 How to use an sc controllerThe probe can be used with all sc controllers. Become familiarized with the controller's func

Page 15 - 3.3 Probe assembly

22OperationSENSOR MENUAN-ISE sc or AISE sc or NISE scCALIBRATEMATRIX CORRMatrix correction options. The most recently used menu is displayed. The cur

Page 16

23OperationSENSOR MENUVALUE POINT 2Enter the values for the 2 point value correction (second point).Note: The following example shows the input when u

Page 17

24OperationSENSOR MENUINFORMATION Information on the matrix correction used per parameterNH4–N 1,2Matrix correction used for ammoniumNO3–N 1,3Matrix c

Page 18

25Operation4.6 Calibration/matrix correctionThe four electrodes with the reference system of the compact sensor cartridge were calibrated with one ano

Page 19

26OperationWhen using the NISE sc probe, only the nitrate membrane and the integrated chloride electrode are active. Cross sensitivities between ammon

Page 20

27OperationThe matrix correction process requires operating steps to be completed on the sc controller and on a LINK2SC-compatible photometer.Refer to

Page 21 - Section 4 Operation

28Operation4.6.4 Performing the matrix correctionNote: Take laboratory value measurements or reference values promptly or, alternatively, take these f

Page 22 - Operation

29Operation4.6.4.2 Value correction 1The one-point value correction VALUE CORR. 1 offers the option of retrospectively performing a matrix correction

Page 23

3Table of contentsSection 1 Technical data...

Page 24

30OperationAll three values form one of the two correction points.4. Look for two correction points where the laboratory values are at least half a de

Page 25

NOTICEMaintenance task 30 days112 monthsNOTICE31Section 5 MaintenanceNote: Do not test the sensor with the usual NH4-N and/or NO3-N standard solutions

Page 26 - 4.6.1 Sensor code calibration

32MaintenanceFigure 14 Chloride electrode 5.3 Replace the sensor cartridgeThe sensor cartridge is replaced as described below and in Figure 15, Page

Page 27

33MaintenanceFigure 15 Replace the sensor cartridge 1 Socket head screw 5 Union nut2 Sensor cartridge 6 Marker hole3 O-ring 7 Socket head wrench4 Sen

Page 28

34Maintenance5.4 StorageTake the probe out of the sample flow and clean thoroughly.Short term storageKeep the membranes and reference system moist (do

Page 29

35Section 6 Troubleshooting6.1 Error messagesIf the sensor is in an error state, the measurement value for this sensor will flash on the display and t

Page 30

36Troubleshooting6.2 WarningsIn the event of a sensor warning, all menus, relays and outputs continue to function as normal but a warning symbol light

Page 31 - Section 5 Maintenance

37Troubleshooting6.3 Troubleshooting 6.3.1 Troubleshooting during operation Symptom Possible cause Corrective measuresIncorrect measurement valuesCali

Page 32 - Maintenance

38Troubleshooting1 applies to AN-ISE sc3 applies to NISE scUnstable measurement valuesAir bubbles, depth of immersionCheck the sensor installationChec

Page 33

Description Catalog number39Section 7 Replacement parts and accessories Description Catalog number Description Catalog number 7.1 Replacement partsAN-

Page 34 - 5.4 Storage

4Section 6 Troubleshooting...356.1 Error

Page 35 - Section 6 Troubleshooting

Description Catalog number40Replacement parts and accessories7.4 Corresponding documentationCleaning unit instruction sheet DOC273.99.90203Rail mount

Page 36 - 6.2 Warnings

41Section 8 Contact HACH Company World HeadquartersP.O. Box 389 Loveland, Colorado 80539-0389 U.S.A. Tel (800) 227-HACH (800) -227-4224 (U.S.A. only)

Page 37 - 6.3 Troubleshooting

42ContactHACH LANGE D.O.O.Fajfarjeva 15 SI-1230 Domžale Tel. +386 (0)59 051 000 Fax +386 (0)59 051 010 [email protected] www.hach-lange.siΗΑCH LANGE

Page 38 - Troubleshooting

43Section 9 Warranty and liabilityThe manufacturer warrants that the supplied product is free of material and manufacturing defects and undertakes to

Page 40

45IndexAAccessories ... 39AISE sc probe ...

Page 42

5Section 1 Technical dataSubject to change.General information AN-ISE sc AISE sc NISE scMeasurement methodPotentiometric measurement using ion-selecti

Page 43

6Technical data1.1 DimensionsFigure 1 Probe dimensions General information about the probeProbe dimensions320 mm × 84.5 mm (12.6 × 3.3 in.) (length ×

Page 44 - Warranty and liability

7Section 2 General information2.1 Safety informationRead the entire operating instructions before the instrument is unpacked, set up or put into opera

Page 45

8General information2.2 General information about the probesThe probes have been developed for use in municipal waste water applications.The ISE probe

Page 46

9General informationFigure 3 Sensor cartridge 2.3 Functional principleIon-selective electrodes have a special membrane to which only a specific type