Technical Specifications
2.4.2 Oxygen sensors (Table 1)
Table 4 Membrane specifications - Oxygen sensors (1)
Membrane Model 2956A 2958A 29552A 2952A
Recommended
applications
Corrosion control,
De-aerated water
Beverage,
Lab. applications
In line wort,
Air/O
2
injection,
Sewage treatment
Corrosion control,
In line beverage,
De-aerated water
Material PFA Tefzel
®
PTFE Tefzel
®
Thickness [µm] 25 12.5 50 25
Calibration gas Air Air Air Air / Pure O
2
Dissolved measurement
range
0 ppb to
20 ppm
0 ppb to
40 ppm
0 ppb to
80 ppm
0 ppb to
80 ppm
Gaseous measurement
range
0 Pa to
50 kPa
0 Pa to
100 kPa
0 Pa to
200 kPa
0 Pa to
200 kPa
Accuracy
The greater of ±1%
of reading
or ± 0.1 ppb
1
,
or ± 1 ppb
2
,
or ± 0.25 Pa
The greater of ±1%
of reading
or ± 1 ppb,
or ± 2 Pa
The greater of ±1%
of reading
or ± 2 ppb,
or ± 5 Pa
The greater of ±1%
of reading
or ± 2 ppb,
or ± 5 Pa
1
Accuracy is ± 0.1 ppb for 410, 510, 362x, 360x and 3655 instruments
2
Accuracy is ± 1 ppb for 366x and 3650 instruments
Integrated radiation dose
limit [rads]
2 x 10
4
10
8
N/A 10
8
Expected current in air @
1 bar 25°C [µA]
26.4 9.4 6.3 5.4
Expected current in pure
O
2
[µA]
132 47 31.4 27
O
2
consumption in O
2
saturated water at 25°C
[µg/hour]
40 14 9.4 8
Temp. compensation
range
– 5 to 60° C
Temp. measuring range – 5 to 100° C
Response time
1
7.2 sec. 9.5 sec. 90 sec. 38 sec.
Recommended min. liquid
flow rate
2
[mL/min]
180 120 50 50
Recommended min. linear
flow rate
2
[cm/sec]
200 100 30 30
Recommended gaseous
flow rate [L/min]
0.1 to 3