5.1.2 Sensor assembly (membrane installation)
Before starting the sensor reassembly, proceed to the sensor maintenance section for anode
and cathode cleaning instructions.
Note: The membrane support is individually machined and paired
with the sensor. For correct sensor operation, it is ESSENTIAL to
keep a membrane support with its respective sensor. Should the
membrane support require replacement, contact your Hach Lange
representative.
Install the membrane support with the groove on the upper
side.
Note: The membrane support is individually machined and paired
with the sensor. Therefore make sure that the correct membrane
support is used on the correct sensor.
Insert the prongs of the
membrane support removal
tool into the membrane
support holes.
Tighten the membrane
support finger tight.
CAUTION
Too much torque will damage the sensor electrodes.
The membrane mounting surface must be clean and even.
Replace the membrane O-ring on the sensor head with a new
one.
Note: The 29039.4 Nitril O-ring can be reused if it is still in good
condition. Membrane O-rings are part of the protection cap kit.