DOC023.52.03231ULTRATURB scbasic/plus/seawaterUser Manual01/2010, Edition 6A© HACH-LANGE GmbH, 2007–2010. All rights reserved. Printed in Germany

Page 2

10General information

Page 3 - Table of contents

11Section 3 InstallationDANGERInstallation may only be carried out by qualified experts in accordance with all local safety regulations.3.1 Assembling

Page 4

12Installation3.1.1 ConnectorsImportant Note: The instrument can be damaged if the connectors and/or the union nuts are attached too tightly. Only han

Page 5 - Section 1 Specifications

Installation133.2 Instrument layoutFig. 3 Instrument layout ULTRATURB sc * Included in the accessory set (connection) LZP8161. Screw top, HRR195 8.

Page 6 - 1.1 Dimensions

14Installation3.3 Connecting sensor cableCAUTION Always lay cables and hoses so that they do not pose a tripping risk.1. Unscrew the protective caps f

Page 7 - Section 2 General information

15Section 4 Operation4.1 Operating the sc controllerThe sensor can be operated with all sc controllers. Prior to using the sensor, familiarise yoursel

Page 8 - 2.4 Handling

16Operation4.4.2 SENSOR SETUPWIPE Initiates a wiping actionVERIFYVerification using CVM moduleVerification with dry standardCALIBRATESTANDARDSelection

Page 9 - 2.6 Function check

Operation17TEST/MAINTPROBE INFOULTRATURBsc Instrument nameEDITED NAMESERIAL NUMBERRANGE 0.001 ... 1000 FNUMODEL NUMBER Item no. sensorSOFTWARE VERS Se

Page 10 - General information

18Operation4.5 Calibration with standard solution1. Open the MAIN MENU. 2. Choose SENSOR SETUP and accept. 3. Choose the related sensor and accept.4.

Page 11 - Section 3 Installation

Operation194.6 Verify using dry standard CVM1. Open the MAIN MENU. 2. Choose SENSOR SETUP and accept. 3. Choose the related sensor and accept.4. Choos

Page 14 - 3.3 Connecting sensor cable

21Section 5 MaintenanceIt has been possible to reduce the scope of the maintenance to be performed by the user to only a few actions. Clearly summaris

Page 15 - Section 4 Operation

22Maintenance9. Remove the screw top and the sealing plug for the measuring chamber and accept OPEN MEAS. CHAMBER.10. Remove the wiper holder and acce

Page 16 - Operation

Maintenance235.4 Replacing desiccantThe ULTRATURB sc turbidity sensor is continuously subject to a wet environment. To provide protection against the

Page 17

24Maintenance5.5 Monitoring test equipmentAs part of a quality assurance system for test equipment monitoring, it is necessary to check the instrument

Page 18

Maintenance25The turbidity of this parent solution is 400, stated in Formazine Attenuation Units (FAU) or Formazine Nephelometric Units (FNU). This so

Page 20

27Section 6 Faults, causes, rectification6.1 Error messagesPossible sensor errors are displayed by the controller.6.2 WarningsPossible warning message

Page 21 - Section 5 Maintenance

28Faults, causes, rectification

Page 22 - Maintenance

29Section 7 Replacement parts and accessories 7.1 Sensor optionsDescription Cat. no.ULTRATURB basic sc turbidity bypass sensor without cable LPV415

Page 23 - 5.4 Replacing desiccant

3Table of contentsSection 1 Specifications ...

Page 24 - 5.5 Monitoring test equipment

30Replacement parts and accessories7.3 AccessoriesDescription Cat. no.Extention cable 1 m (3.3 ft.) 6122400Extention cable 5 m (16.40 ft.) LZX848Exten

Page 25

31Section 8 Warranty, liability and complaintsThe manufacturer warrants that the product supplied is free of material and manufacturing defects and un

Page 26

32Warranty, liability and complaints

Page 27 - 6.2 Warnings

33Section 9 Contact HACH CompanyWorld HeadquartersP.O. Box 389Loveland, Colorado80539-0389 U.S.A.Tel (800) 227-HACH(800) -227-4224(U.S.A. only)Fax (9

Page 28 - Faults, causes, rectification

34ContactHACH LANGE D.O.O.Fajfarjeva 15SI-1230 DomžaleTel. +386 (0)59 051 000Fax +386 (0)59 051 [email protected]ΗΑCH LANGE E.Π.Ε

Page 29 - 7.2 Replacement parts

35Appendix A ModBUS Register InformationTable 5: Sensor ModBUS RegistersGroup Name Register # Data Type Length R/W Descriptionmeasurement 40001 Float

Page 30 - 7.3 Accessories

36ModBUS Register Informationuser_cal_int 40082 Integer 1 R/W user cal intcal_out_cfg 40083 Integer 1 R/W cal. Output mode meas_mode 40084 Integer 1 R

Page 31

4Table of contents

Page 32

5Section 1 Specifications Specifications are subject to change without notice.ComponentsMicroprocessor-controlled turbidity bypass sensor ULTRATURB sc

Page 33 - Section 9 Contact

6Specifications1.1 DimensionsFig. 1 ULTRATURB sc dimensions105.15 mm[4.14 in]221 mm[8.7 in]216 mm[8.5 in]233,4 mm[9.19 in]111 mm[4.37 in]85 mm[3.35 in

Page 34

7Section 2 General information2.1 Safety informationPlease read this entire manual before unpacking, setting up or operating this equipment. Pay atten

Page 35

8General information2.2 ApplicationsULTRATURB sc bypass sensors are innovative precision turbidity measuring instruments developed using the latest te

Page 36 - ModBUS Register Information

General information92.5 Items supplied• ULTRATURB sc sensor• Connecting cable (length as per order)• User Manual• Factory test certificate• Accessory