User Manual
DOC022.52
.80029
Ammonium Probe: Model ISENH418101 or ISENH418103
Safety information
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed. A symbol on the instrument is referenced in the
manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European
electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to
the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life
equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
Specifications
Note: Specifications are subject to change without notice.
Specifications Details
Probe type Digital combination probe with a non-refillable reference junction and a built-
in temperature sensor
Range 0.018 mg/L (10
-6
M) to 9,000 mg/L (0.5 M) as Ammonium
Sample pH range pH 0 to 8.5
Linear region 0.9 mg/L to 9,000 mg/L as Ammonium
Slope -58 mV/decade (90 to 110% at 25 °C (77 °F) in linear range per Nernstian
theoretical value)
Operating temperature range 0 to 50 °C (32 to 122 °F)
Storage temperature range 5 to 35 °C (41 to 95 °F)
Junction Double junction (ceramic porous pin and annular porous Teflon
®
)
Reference type Ag/AgCl
Fill solution driTEK Gel (non-refillable)
Response time in linear region < 60 seconds (application dependent)
Minimum sample volume 25 mL
Minimum immersion depth 25.4 mm (1 in.)
Dimensions Diameter: 12 mm (0.47 in.)
Length: 175 mm (6.89 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)
Cable connection M12 digital output and connector compatible with HQd meters
Product overview
The ISENH418101 or ISENH418103
probe is a combination ammonium probe with a
built-in temperature sensor (Figure 1). The probe is available with a 1 or 3 m (3.28 or
9.84 ft) cable and is intended for laboratory use. The probe measures ammonium
concentration in water samples.
1
Vue de la page 0
1 2 3 4 5 6 ... 13 14

Résumé du contenu

Page 1 - User Manual

User ManualDOC022.52.80029Ammonium Probe: Model ISENH418101 or ISENH418103Safety informationPrecautionary labelsRead all labels and tags attached to t

Page 2 - Calibration

7. If prompted, enter a name for the new method settings. Additional changes made tothe settings of an existing method are automatically saved with th

Page 3

Message or symptom Possible cause ActionSlow response timeDirty sensing element Clean the probe (refer to Clean the probeon page 10).Low sample temper

Page 4 - Measurement—direct method

Message or symptom Possible cause ActionOut of limitsCheck standard value is outside oflimits set in the current methodMake sure that the standard is

Page 5 - Low-level measurements

Percent recoveryUse the following formula to calculate the percent recovery when the sample volume is25 mL:E = (C x V1) / V2R = (A / (E + S)) x 100 •S

Page 6 - Run a check standard

HACH COMPANY World HeadquartersP.O. Box 389, Loveland, CO 80539-0389 U.S.A.Tel. (970) 669-3050(800) 227-4224 (U.S.A. only)Fax (970) 669-2932orders@hac

Page 7 - Advanced operation

Figure 1 Probe overview1 Sensor protection cap 3 Sensing element2 Reference junction 4 1 or 3 m (3.28 or 9.84 ft) cablePreparation for usePre

Page 8 - Change calibration options

• The calibration is recorded in the electrode and the data log. The calibration is alsosent to a PC, printer or flash memory stick if connected.•Air

Page 9 - Change check standard options

11. Push Store toaccept thecalibration andreturn to themeasurementmode.Measurement—direct methodBefore measurement:The probe must have the correct se

Page 10 - Troubleshooting

1. Connect theprobe to themeter. Make surethat the cablelocking nut issecurelyconnected to themeter. Turn themeter on.2. Prepare aminimum of 25 mLof

Page 11

• The sample pH is in the range that is given in the specifications.•Omission of the ISA is accepted by the regulatory reporting agency (if the measur

Page 12

the acceptance criteria is set to Cal Expires on Failure: Yes, the display shows thecalibration icon and a question mark until the probe is recalibrat

Page 13

Option DescriptionUnits Sets the preferred unit for ISE measurements—mg/L (default), µg/L, g/L,g/kg, mol/L, mmol/L, mol/kg, %, ppm or ppb.Note: The mV

Page 14 - 05/2013, Edition 4

Option DescriptionStd Set Values When Std Set is set to Custom, sets the standard set values.Up to five standard values can be made. Each standard val