User Manual
DOC022.52
.80030
Chloride Probe: Model ISECI18101 or ISECI18103
Safety information
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed. A symbol on the instrument is referenced in the
manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European
electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to
the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life
equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
Specifications
Note: Specifications are subject to change without notice.
Specifications Details
Probe type Digital combination probe with a non-refillable reference junction and a built-
in temperature sensor
Range 0.1 mg/L (3x10
-6
M) to 35,500 mg/L (1 M) Chloride
Sample pH range pH 1 to 12
Linear region 7 mg/L to 35,500 mg/L
Slope 58 mV/decade (90 to 110% at 25 °C (77 °F) in linear range per Nernstian
theoretical value)
Operating temperature range 5 to 50 °C (41 to 122 °F)
Storage temperature range 5 to 35 °C (41 to 95 °F)
Junction Double junction (ceramic porous pin and annular porous Teflon
®
)
Reference type Ag/AgCl
Fill solution driTEK Gel (non-refillable)
Stabilization time in linear region < 60 seconds (application dependent)
Minimum sample volume 25 mL
Minimum immersion depth 25.4 mm (1 in.)
Dimensions Diameter: 12 mm (0.47 in.)
Length: 175 mm (6.89 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)
Cable connection M12 digital output and connector compatible with HQd meters
Product overview
The ISECl18101 or ISECl18103 probe is a combination
chloride probe with a built-in
temperature sensor (Figure 1). The probe is available with a 1 or 3 m (3.28 or 9.84 ft)
cable and is intended for laboratory use. The probe measures chloride concentration in
water samples.
1
Vue de la page 0
1 2 3 4 5 6 ... 13 14

Résumé du contenu

Page 1 - User Manual

User ManualDOC022.52.80030Chloride Probe: Model ISECI18101 or ISECI18103Safety informationPrecautionary labelsRead all labels and tags attached to the

Page 2 - Calibration

MaintenanceClean the probeClean the probe when:•Drifting/inaccurate readings occur as a result of contamination on the sensingelement or improper stor

Page 3

Message or symptom Possible cause ActionSlope out of range (refer to Check probe responseon page 12)Ionic strength adjustor (ISA) notusedAdd ISA to ea

Page 4 - Measurement—direct method

Message or symptom Possible cause ActionTemperature out of rangeCalibration temperature value isoutside of rangeMake sure that the sample temperature

Page 5 - Low-level measurements

• V1 = spike volume (mL)•V2 = spike volume (mL) + 25 mL sample volume• E = expected concentration of spike (mg/L)• R = percent recovery• A = actual re

Page 6 - Run a check standard

HACH COMPANY World HeadquartersP.O. Box 389, Loveland, CO 80539-0389 U.S.A.Tel. (970) 669-3050(800) 227-4224 (U.S.A. only)Fax (970) 669-2932orders@hac

Page 7 - Advanced operation

Figure 1 Probe overview1 Sensor protection cap 3 Sensing element2 Reference junction 4 1 or 3 m (3.28 or 9.84 ft) cablePreparation for usePre

Page 8 - Change calibration options

• The calibration is recorded in the electrode and the data log. The calibration is alsosent to a PC, printer or flash memory stick if connected.•Air

Page 9 - Change check standard options

11. Push Store toaccept thecalibration andreturn to themeasurementmode.Measurement—direct methodBefore measurement:The probe must have the correct se

Page 10 - Troubleshooting

1. Connect theprobe to themeter. Make surethat the cablelocking nut issecurelyconnected to themeter. Turn themeter on.2. Prepare aminimum of 25 mLof

Page 11

• The sample pH is in the range that is given in the specifications.•Omission of the ISA is accepted by the regulatory reporting agency (if the measur

Page 12

5. If the display shows Check Standard Passed, the check standard measurement iswithin the accepted limits set by the administrative user. Select Done

Page 13

4. Select Measurement Options and update the settings:Option DescriptionUnits Sets the preferred unit for ISE measurements—mg/L (default), µg/L, g/L,g

Page 14 - 05/2013, Edition 4

Option DescriptionMinimum CalPointsSets the minimum number of calibration points necessary before acalibration can be completed—2 or 3.Slope Limit Set