DOC023.52.93069Polymetron 9586 OxygenScavenger Analyzer11/2013, Edition 3User Manual

Page 2

Analyzer mountingC A U T I O N Personal injury hazard. Instruments or components are heavy. Use assistance to install or move. Make sure thatthe wall

Page 3

Wiring overviewFigure 2 shows an overview of the wiring connections inside the controller with the high voltagebarrier removed. The left side of the f

Page 4 - Table of Contents

W A R N I N G Potential Electrocution Hazard. If this equipment is used outdoors or in potentially wet locations, aGround Fault Interrupt device must

Page 5 - Specifications

Table 2 DC power wiring information (DC powered models only)Terminal Description Color—North America Color—EU1 +24 VDC Red Red2 24 VDC return Blac

Page 6 - General information

Alarms and relaysThe controller is equipped with four unpowered, single pole relays rated 100-250 VAC, 50/60 Hz,5 amp resistive maximum. Contacts are

Page 7

W A R N I N G Potential electrocution hazard. AC mains powered controllers (115 V–230 V) are designed for relayconnections to AC mains circuits (i.e.,

Page 8 - Product components

Analog output connectionsW A R N I N G Potential Electrocution Hazard. Always disconnect power to the instrument when making electricalconnections.W A

Page 9 - Installation

4. Make connections with twisted-pair shielded wire and connect the shield at the controlledcomponent end or at the control loop end.• Do not connect

Page 10 - N O T I C E

Figure 4 Discrete input wiring connectionsTable 4 Input connections Discrete inputs Connector position - Switch input Connector position - Voltagein

Page 11 - W A R N I N G

Note: In switch input mode the controller supplies 12 volts to the switch and is not isolated from the controller. Involtage input mode the inputs are

Page 13 - English 11

Install the reagentsW A R N I N G Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protectiveequipment appropr

Page 14

Figure 7 Keypad and front panel overview1 Instrument display 5 BACK key. Moves back one level in the menustructure.2 Cover for secure digital m

Page 15

Figure 8 Example of Main Measurement screen1 Home screen icon 7 Warning status bar2 Sensor name 8 Date3 SD Memory card icon 9 Analog outp

Page 16

1. From the graphical display screen use the up and down arrow keys to select a graph and pushthe HOME key.2. Select an option:Option DescriptionMEASU

Page 17

Option DescriptionDISPLAY SETUP Configures the controller display.ADJUST ORDER—View and modify the measurement display order.• SEE CURRENT ORDER—View

Page 18

2. Select an option and push enter to activate the menu item.Using the secure digital memory (SD) cardAn SD card must be installed in the controller.•

Page 19

3. Select the time period from which logs are to be saved.Option DescriptionLAST DAY All logs from the last full 24 hours, starting from 12:00 a.m., a

Page 20 - User interface and navigation

Firmware updates with SD cardsThe latest firmware updates can be placed on an SD card. The SD card can then be used to updatethe controller or device

Page 21

d. Push enter to return to the MANAGE CONFIGURATION menu.4. To copy settings from the SD card to a controller (or a device connected to it):a. Select

Page 22 - Graphical display

5. The raw temperature value is displayed. Push enter.6. Enter the correct value if different from that displayed and push enter.7. Push enter to conf

Page 23 - System startup

Table of ContentsAdditional information...3Specifications...

Page 24 - 22 English

4. Select the option for the output signal during calibration:Option DescriptionACTIVE The instrument sends the current measured output value during t

Page 25 - Updating software

3. Select CAL OPTIONS and push enter.4. Use the arrow keys to select an option and push enter.Option DescriptionCAL REMINDER Sets a reminder for the n

Page 26

Table 6 Controller (continued)Problem ResolutionNo relay activationMake sure relay connections are secure.If using an external power source, make sur

Page 27 - Backup settings to an SD card

Table 6 Controller (continued)Problem ResolutionController will not power up, or powers upintermittentlyMake sure the AC power connections are proper

Page 28 - Operation

Option DescriptionOUTPUT CAL• OUTPUT 1 • OUTPUT 2 Lets the user calibrate the 4–20 mA outputs with a 250 ohm resistor in series tothe mA output termin

Page 29 - Zero calibration

Table 8 Sensor DIAG/TEST menuOption DescriptionMODULE INFORMATION Shows information about the sensor module.SENSOR INFORMATION Shows the name and ser

Page 30 - Change calibration options

Table 9 Warning list for oxygen scavenger sensors (continued)Warning Description ResolutionTEMP TOO LOW The measured temperature is <0 °CIncrease

Page 31 - Troubleshooting

Spare parts kit for 2 years - 09186=A=8000 (continued)Description Quantity Item no.O-Ring for Venturi injection nozzle 1 356099,05090 Plastic cleanin

Page 32 - 30 English

2. COMPOSITION / INFORMATION ON INGREDIENTSCAS no: 108-18-9 Molecular weight: 101.19Molecular formula: C6H15NEC index number: 612-048-00-5EINECS: 203-

Page 33 - Test and maintenance menu

9. PHYSICAL AND CHEMICAL PROPERTIESpH value: not availableMelting temperature: -96°CBoiling temperature: 83 - 84°CIgnition temperature: 295°C DIN51794

Page 34

Temperature calibration... 26Zero calibration...

Page 35 - Warning and error conditions

14. TRANSPORT INFORMATIONDOT and IATA:• Shipping Name: Diisopropylamine• UN Number: UN1158• Hazard Class: 3.8• Packing Group: PG II• Label(s): Flammab

Page 37 - MSDS - Diisopropylamine

HACH COMPANY World HeadquartersP.O. Box 389, Loveland, CO 80539-0389 U.S.A.Tel. (970) 669-3050(800) 227-4224 (U.S.A. only)Fax (970) 669-2932orders@hac

Page 38 - 36 English

Additional informationAdditional information is available on the manufacturer's website.SpecificationsSpecifications are subject to change withou

Page 39 - English 37

Specification DetailsPower requirements AC powered controller: 100-240 VAC ±10%, 50/60 Hz; Power 50 VA with 7 Wsensor/network module load, 100 VA with

Page 40 - 38 English

Make sure that the protection provided by this equipment is not impaired. Do not use or install thisequipment in any manner other than that specified

Page 41

EMC compliance statement (Korea)Type of equipment Additional informationA 급 기기( 업무용 방송통신기자재 )이 기기는 업무용 (A 급 ) 전자파적합기기로서 판매자 또는 사용자는 이 점을 주의하시기 바라며, 가정

Page 42 - Hach Company, 2013

Product overviewThe analyzer is designed to continuously measure the amount of oxygen scavengers (hydrazine orcarbohydrazide) in water.Figure 1 Front