DOC023.52.93069Polymetron 9586 OxygenScavenger Analyzer11/2013, Edition 3User Manual
Analyzer mountingC A U T I O N Personal injury hazard. Instruments or components are heavy. Use assistance to install or move. Make sure thatthe wall
Wiring overviewFigure 2 shows an overview of the wiring connections inside the controller with the high voltagebarrier removed. The left side of the f
W A R N I N G Potential Electrocution Hazard. If this equipment is used outdoors or in potentially wet locations, aGround Fault Interrupt device must
Table 2 DC power wiring information (DC powered models only)Terminal Description Color—North America Color—EU1 +24 VDC Red Red2 24 VDC return Blac
Alarms and relaysThe controller is equipped with four unpowered, single pole relays rated 100-250 VAC, 50/60 Hz,5 amp resistive maximum. Contacts are
W A R N I N G Potential electrocution hazard. AC mains powered controllers (115 V–230 V) are designed for relayconnections to AC mains circuits (i.e.,
Analog output connectionsW A R N I N G Potential Electrocution Hazard. Always disconnect power to the instrument when making electricalconnections.W A
4. Make connections with twisted-pair shielded wire and connect the shield at the controlledcomponent end or at the control loop end.• Do not connect
Figure 4 Discrete input wiring connectionsTable 4 Input connections Discrete inputs Connector position - Switch input Connector position - Voltagein
Note: In switch input mode the controller supplies 12 volts to the switch and is not isolated from the controller. Involtage input mode the inputs are
Install the reagentsW A R N I N G Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protectiveequipment appropr
Figure 7 Keypad and front panel overview1 Instrument display 5 BACK key. Moves back one level in the menustructure.2 Cover for secure digital m
Figure 8 Example of Main Measurement screen1 Home screen icon 7 Warning status bar2 Sensor name 8 Date3 SD Memory card icon 9 Analog outp
1. From the graphical display screen use the up and down arrow keys to select a graph and pushthe HOME key.2. Select an option:Option DescriptionMEASU
Option DescriptionDISPLAY SETUP Configures the controller display.ADJUST ORDER—View and modify the measurement display order.• SEE CURRENT ORDER—View
2. Select an option and push enter to activate the menu item.Using the secure digital memory (SD) cardAn SD card must be installed in the controller.•
3. Select the time period from which logs are to be saved.Option DescriptionLAST DAY All logs from the last full 24 hours, starting from 12:00 a.m., a
Firmware updates with SD cardsThe latest firmware updates can be placed on an SD card. The SD card can then be used to updatethe controller or device
d. Push enter to return to the MANAGE CONFIGURATION menu.4. To copy settings from the SD card to a controller (or a device connected to it):a. Select
5. The raw temperature value is displayed. Push enter.6. Enter the correct value if different from that displayed and push enter.7. Push enter to conf
Table of ContentsAdditional information...3Specifications...
4. Select the option for the output signal during calibration:Option DescriptionACTIVE The instrument sends the current measured output value during t
3. Select CAL OPTIONS and push enter.4. Use the arrow keys to select an option and push enter.Option DescriptionCAL REMINDER Sets a reminder for the n
Table 6 Controller (continued)Problem ResolutionNo relay activationMake sure relay connections are secure.If using an external power source, make sur
Table 6 Controller (continued)Problem ResolutionController will not power up, or powers upintermittentlyMake sure the AC power connections are proper
Option DescriptionOUTPUT CAL• OUTPUT 1 • OUTPUT 2 Lets the user calibrate the 4–20 mA outputs with a 250 ohm resistor in series tothe mA output termin
Table 8 Sensor DIAG/TEST menuOption DescriptionMODULE INFORMATION Shows information about the sensor module.SENSOR INFORMATION Shows the name and ser
Table 9 Warning list for oxygen scavenger sensors (continued)Warning Description ResolutionTEMP TOO LOW The measured temperature is <0 °CIncrease
Spare parts kit for 2 years - 09186=A=8000 (continued)Description Quantity Item no.O-Ring for Venturi injection nozzle 1 356099,05090 Plastic cleanin
2. COMPOSITION / INFORMATION ON INGREDIENTSCAS no: 108-18-9 Molecular weight: 101.19Molecular formula: C6H15NEC index number: 612-048-00-5EINECS: 203-
9. PHYSICAL AND CHEMICAL PROPERTIESpH value: not availableMelting temperature: -96°CBoiling temperature: 83 - 84°CIgnition temperature: 295°C DIN51794
Temperature calibration... 26Zero calibration...
14. TRANSPORT INFORMATIONDOT and IATA:• Shipping Name: Diisopropylamine• UN Number: UN1158• Hazard Class: 3.8• Packing Group: PG II• Label(s): Flammab
HACH COMPANY World HeadquartersP.O. Box 389, Loveland, CO 80539-0389 U.S.A.Tel. (970) 669-3050(800) 227-4224 (U.S.A. only)Fax (970) 669-2932orders@hac
Additional informationAdditional information is available on the manufacturer's website.SpecificationsSpecifications are subject to change withou
Specification DetailsPower requirements AC powered controller: 100-240 VAC ±10%, 50/60 Hz; Power 50 VA with 7 Wsensor/network module load, 100 VA with
Make sure that the protection provided by this equipment is not impaired. Do not use or install thisequipment in any manner other than that specified
EMC compliance statement (Korea)Type of equipment Additional informationA 급 기기( 업무용 방송통신기자재 )이 기기는 업무용 (A 급 ) 전자파적합기기로서 판매자 또는 사용자는 이 점을 주의하시기 바라며, 가정
Product overviewThe analyzer is designed to continuously measure the amount of oxygen scavengers (hydrazine orcarbohydrazide) in water.Figure 1 Front