Hach-lange ORBISPHERE K1100_M1100 Operator Manual Manuel d'utilisateur Page 24

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 114
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 23
22
Installation
3.2 Installation check list
For a complete installation, proceed to the following actions following the instructions in this
manual very carefully:
1. Install the 410 analyzer (section 3.3 on page 23 through section 3.8 on page 34).
2. Install the sensor in the sample pipe or flow chamber and connect it to the instrument
(section 3.9 on page 35).
3. Turn the instrument on and set the operating language (section 13.2 on page 95).
4. Set the security levels, users ID’s and passwords (Section 10 on page 87).
5. Verify that the sensor cap selected (i.e. low or high range: K/M1100-L or K/M1100-H)
corresponds to the sensor in use (section 6.2 on page 52).
6. The measurement liquid is set by default to water for the K1100 and M1100 high range
sensors, and beer for the M1100 low range sensor. For the M1100 low range sensor,
change this to water if the sample being measured is not beer (section 6.2 on page 52).
7. Enable thermal cutoff and set it to the minimum possible value (e.g. 10°C if the liquid in the
sample line is at 0-5°C). This is critical to ensure long-term stability (section 6.2 on
page 52).
8. Set the measurement interval according to your response time needs (section 6.2 on
page 52).
9. Enable the calibration timer and set the number of days to 180 days (6 months) or
according to your quality needs (section 13.2 on page 95), to ensure a regular sensor
calibration.
10. Enable the service timer and set the number of days to 360 days (1 year) or according to
your quality needs (Service timer on page 95), to ensure a regular sensor service.
11. If required, program the relay and analog outputs to ensure an event is set when a
calibration or service is required. These can be programmed for each as custom events
(Section 8 on page 61). Note that when a calibration or service timer has expired, the
instrument will automatically display this information as a warning message (see the table
entitled List of Events on page 101).
As the system is factory calibrated and does not require an initial calibration (refer to Calibration
Menu on page 55 for more details), the instrument and sensor should now be ready for
operation. If a problem should arise, please refer initially to Troubleshooting on page 99. If the
difficulty cannot be overcome, please contact your Hach Lange representative who will be
happy to assist you.
Vue de la page 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 113 114

Commentaires sur ces manuels

Pas de commentaire